SOBOL 1NM










SOBOL 1NM

FUNCTIONALITY


The device is used for treatment of patients with ominous bradycardia caused by sick sinus syndrome, AV blocks and other diseases. 

APPLICATION

• Cardiovascular care and cardiac surgery units. 
• Intensive care units.
• Cardiological emergency teams.
• Electrophysiological intracardiac studies laboratories.
• Air medical services.

TECHNICAL SPECIFICATIONS

• The device provides pacing in AAO, AAI, VVO, VVI modes under the universal code of nomenclature NGB-NASPE/BPEG.
• Stimulation pulse parameters:
• Pulse amplitude - 1-20mA.
• Pulse amplitude setting resolution - 1mA.
• Pulse length - 0.5-4 msec.
• Pulse length setting resolution - 0.5msec.
• Pulse rate - 40-200 ppm.
• Pulse rate setting resolution - 10ppm.
• Power supply: batteries of 6F22 type (“Krona”).
• Safety requirements:
• Electrocution protection CF under Federal standard 50267.0-92 (IEC 601-I-88)
• Dimensions - at most 170x90x32 mm.
• Weight - at most 300 g.

ADDITIONAL OPTIONS

• Entry into “emergency option” with pressing a button. 
• Protection against random parametric variation.